| Intensity: | 0-100mA (500Ω load) adjustable | Pro Function 3: | Patient Lock/unlock of settings |
| Output: | Constant current |
Channels: |
Two |
| Pulse Rate: | 1-200Hz - 50 selectable frequencies |
Waveform: |
Symmetrical bi-phasic rectangular |
| Pulse Width (µs): | 25, 50, 75, 100, 125, 150, 175, 200, 225 & 250 |
Display: |
Modulation modes - changes visible on display |
| 5 Modes: | Constant, Burst (I + II), Modulation (PW or PR) |
Timer (mins): |
continuous, 10, 15, 20, 30, 45, 60 |
| Burst I: |
2 fixed bursts per sec. Each burst consists of 7 pulses. Pulse Width adjustable |
Memory: |
Saves setting for next time the unit is used. Memory erased when batteries changed |
| Burst II: |
1 fixed burst per 2 secs. Each burst consists of 80 pulses. Pulse Width and Rate adjustable |
Load detector: |
If an electrode accidentally detaches from the body or leadwire, intensity drops to zero |
| Width Modulation: |
Pulse Width modulates from 25 - 250µs, ramps up 8 secs & ramps down 8 secs. Pulse Rate is adjustable |
Auto Shutdown: |
If intensity is 0 on both channels and the unit is not used for 5 mins, it automatically shuts down |
| Rate Modulation: |
Pulse Rate modulates decreasing from 100Hz - 20Hz over 8 secs. It then increases from 20Hz - 100Hz over 8 secs. The pattern is then repeated |
Power Source: | 2 x 1.5v AA batteries (Type LR6) |
| Battery Life: |
100 hours approx |
| Low Battery: | Flashing symbol indicates low battery |
| Mode Change: | When mode is changed, intensity drops to 0, then gradually increasing to the level set by user |
Weight / Size: | 100g, 10.5 x 6.5 x 2.75cm |
| Output Jack: | CE Touchproof |
| Pro Function 1: | Memorizes up to 99 hours used time | Attached Belt Clip |
| Pro Function 2: | Memorises times used (if used for more than 5 mins + intensity activated) |